Internal Jugular Vein as Alternative Access for Implantation of a Wireless Pulmonary Artery Pressure Sensor.
Publication Title
Circ Heart Fail
Document Type
Article
Publication Date
8-1-2019
Keywords
cards
Abstract
BACKGROUND: A wireless pulmonary artery pressure sensor (CardioMEMS) is approved for implantation via the femoral vein. The internal jugular vein (IJ) is an attractive alternative access route commonly used in pulmonary artery catheterization.
METHODS AND RESULTS: Retrospective chart review was performed for all sensor implants from 10 providers at 4 centers from September 2016 to June 2018. To compare procedural outcomes and discharge efficiency between groups, multivariate analyses incorporating potential confounders were performed. Seventy-three (28%) patients had femoral access, and 189 (72%) had IJ access; demographics were similar between the groups. Complications, including one case of hematoma and 4 cases of mild hemoptysis, and 30-day mortality (2%-3%) did not differ between groups. Provider preference for IJ access substantially increased over time, with IJ accounting for 90% of cases in 2018. After risk-adjustment, IJ cases had 20% (5%-33%) shorter fluoroscopy time (
CONCLUSIONS: IJ access for CardioMEMS implant is a safe alternative associated with superior procedural and discharge outcomes. Implanters at 4 high-volume centers adopted IJ access as the preferred implant approach.
Area of Special Interest
Cardiovascular (Heart)
Specialty/Research Institute
Cardiology
Specialty/Research Institute
Center for Cardiovascular Analytics, Research + Data Science (CARDS)