Mid-Term Outcomes of Transcatheter Aortic Valve Replacement in Extremely Large Annuli With Edwards SAPIEN 3 Valve.
Publication Title
JACC Cardiovasc Interv
Document Type
Article
Publication Date
1-27-2020
Abstract
OBJECTIVES: The aim of this study was to report the 1-year results of transcatheter aortic valve replacement (TAVR) with the Edwards SAPIEN 3 (S3) valve in extremely large annuli.
BACKGROUND: Favorable 30-day outcomes of S3 TAVR in annuli >683 mm
METHODS: From December 2013 to December 2018, 105 patients across 15 centers with mean area 721.3 ± 36.1 mm
RESULTS: The mean age was 76.9 ± 10.4 years, and Society of Thoracic Surgeons predicted risk score averaged 5.2 ± 3.4%. One-year overall mortality and stroke rates were 18.2% and 2.4%, respectively. Quality-of-life index improved from baseline to 30 days and at 1 year (p < 0.001 for both). Mild paravalvular aortic regurgitation occurred in 21.7% of patients, while moderate or greater paravalvular aortic regurgitation occurred in 4.3%. Mild and moderate or severe transvalvular aortic regurgitation occurred in 11.6% and 0%, respectively. Valve gradients remained stable at 1 year.
CONCLUSIONS: S3 TAVR in annular areas >683 mm
Area of Special Interest
Cardiovascular (Heart)
Specialty/Research Institute
Cardiology